Skip to main content
1019398 Residual Chlorine Sensor

1019398 ProMinent Bulgaria Sales

If you are looking for a reliable supplier where you can purchase the 1019398 Residual Chlorine Sensor, you are in the right place. IMPEXRON Ltd. sells in the Bulgaria marketProMinent all product groups belonging to the brand, with fast delivery times and competitive prices.

1019398 Residual Chlorine Sensor
Product1019398
BrandProMinent
HS Code90278930
CountryBulgaria
Request a price
Delivery time: Determined at the time of quotation
VISA Mastercard Bank Transfer

Technical Specification

CLE 3.1 mA-5ppm sensor sensor for measuring free chlorine in clear water, with increased selectivity with respect to bound chlorine. Applications include applications in drinking water with higher Proportions of bound chlorine and in swimming pools for Determination of bound chlorine from the total chlorine difference minus free chlorine. Measurement variable: free chlorine Measurement range: 0.05... 5.0 mg/l Reference method: DPD1 pH range: 5.5... 8.0 Temperature: 5... 45°C max. pressure: 1.0 bar Inflow: 30... 60 l/h (in bypass, open outlet of the measurement water) Output signal: 4... 20 mA ~ measurement range temperature compensated, uncalibrated, no galvanic isolation Sensor fitting: DGM or DLG Measurement principle: amperometric, 2 electrodes, membrane-covered Installation in the bypass to the process water line Connection to the controller via control line LiYY 2x0.25 qmm Order number: 725122 (sold by the meter) (Test lead not included) Note: **************** For installation in the DLG III pass-through sensor, a Assembly kit order number 815079 required. (not included)